Pages

THE HISTORY OF CIVIL ENGINEER

Showing posts with label Calibration of Temperature Measuring Device. Show all posts
Showing posts with label Calibration of Temperature Measuring Device. Show all posts

Monday, November 9, 2020

Calibration of Temperature Measuring Device (ASTM Standard)


 Apparatus:

4.1 Container, shall be made of no absorptive material and large enough to provide at least 3 in {75 mm} of concrete in all directions around the sensor of the temperature measuring device; concrete cover must also be at least three times the nominal maximum size of the coarse aggregate.

4.2 Temperature Measuring Device, shall be capable of accurately measuring the temperature of the freshly mixed concrete to ± 1°F [±0.5°C] throughout a range of 30°to120°F [0° to 50°C]. The temperature measuring device shall require immersion of 3 in {75 mm} or less during operation.

4.3 Partial immersion liquid-in-glass thermometers (and possibly other types) shall have a permanent mark to which the device must be immersed without applying a correction factor.

4.4 Reference Temperature Measuring Device, shall be readable and accurate to ±0.5°F [0.2°C] at the verification points in 5.1. A certificate or report that verifies the accuracy shall be available in the laboratory for review. The certificate or report shall provide documentation that the reference standard used in the verification is traceable to NIST.

5. Calibration of Temperature Measuring Device   

5.1 Each temperature measuring device used for determining temperature of freshly mixed concrete shall be calibrated annually, or whenever there is a question of accuracy. This calibration shall be performed by comparing the readings of the temperature measuring device at two temperatures at least 30°F [15°C] apart.

5.2 Calibration of the temperature measuring device may be made in oil or other suitable baths having uniform density if provision is made to:

5.2.1 Maintain the bath temperature constant within 0.5°F [0.2°C] during the period of the test.

5.2.2 Have both the temperature and reference temperature measuring device maintained in the both for a minimum of 5 min before reading temperatures.

5.2.3 Continuously circulate the bath liquid to provide a uniform temperature.

5.2.4 Slightly tap thermometers containing liquid to avoid adhesion of the liquid to the glass if the temperature exposure is being reduced.

Apparatus:

4.1 Container, shall be made of no absorptive material and large enough to provide at least 3 in {75 mm} of concrete in all directions around the sensor of the temperature measuring device; concrete cover must also be at least three times the nominal maximum size of the coarse aggregate.

4.2 Temperature Measuring Device, shall be capable of accurately measuring the temperature of the freshly mixed concrete to ± 1°F [±0.5°C] throughout a range of 30°to120°F [0° to 50°C]. The temperature measuring device shall require immersion of 3 in {75 mm} or less during operation.

4.3 Partial immersion liquid-in-glass thermometers (and possibly other types) shall have a permanent mark to which the device must be immersed without applying a correction factor.

4.4 Reference Temperature Measuring Device, shall be readable and accurate to ±0.5°F [0.2°C] at the verification points in 5.1. A certificate or report that verifies the accuracy shall be available in the laboratory for review. The certificate or report shall provide documentation that the reference standard used in the verification is traceable to NIST.

5. Calibration of Temperature Measuring Device   

5.1 Each temperature measuring device used for determining temperature of freshly mixed concrete shall be calibrated annually, or whenever there is a question of accuracy. This calibration shall be performed by comparing the readings of the temperature measuring device at two temperatures at least 30°F [15°C] apart.

5.2 Calibration of the temperature measuring device may be made in oil or other suitable baths having uniform density if provision is made to:

5.2.1 Maintain the bath temperature constant within 0.5°F [0.2°C] during the period of the test.

5.2.2 Have both the temperature and reference temperature measuring device maintained in the both for a minimum of 5 min before reading temperatures.

5.2.3 Continuously circulate the bath liquid to provide a uniform temperature.

5.2.4 Slightly tap thermometers containing liquid to avoid adhesion of the liquid to the glass if the temperature exposure is being reduced.

WhatsApp on Friday delayed a data-sharing

WhatsApp on Friday delayed a data-sharing change as users worried about privacy fled the Facebook-owned messaging service and flocked to riv...